Chemistry · Solutions

Beer’s Law

Essential Question

How can color tell us concentration?

In the last lesson you learned how to dilute a solution to any molarity you need. Now you will use those dilutions to find the concentration of a solution just by shining light through it. By the end of this lesson, you should be able to explain why a solution has a color, describe how a spectrophotometer measures transmittance and absorbance, use Beer’s law (A = εLc), and use a standard curve to find the concentration of an unknown.

Video not loading? Watch it on YouTube.

Why Solutions Have Color

  • White light has all colors
  • A solution absorbs some colors
  • We see the colors that pass through
Rainbow-colored arrows labeled white light shine into a cuvette of pink cobalt chloride solution. Only pink arrows come out the other side, labeled pink light. A note says CoCl2(aq) absorbs blue-green light.

White light is a mix of every color. When it shines through a colored solution, the dissolved particles absorb some of those colors and let the rest pass through. The color you see is the light that makes it out the other side. A cobalt(II) chloride solution, CoCl2(aq), looks pink because it absorbs blue-green light and lets the reds and pinks through.

Darker Means More Concentrated

  • More solute particles in the light path
  • More particles absorb more light
  • Darker color = higher concentration
Three cuvettes of pink solution labeled low c, medium c, and high c. The darker cuvettes hold more particles. An arrow underneath is labeled more light absorbed.

Each solute particle in the path of the light has a chance to absorb some of it. A more concentrated solution has more particles packed into the same space, so it absorbs more light and looks darker. That link between how dark a solution is and how much solute it holds is the whole idea behind Beer’s law. Our eyes can only compare colors roughly, so chemists use an instrument to measure the light exactly.

The Spectrophotometer

  • Picks one wavelength (color) of light
  • Detector measures the light that gets through
  • Reports %T or absorbance (A)
Four boxes connected by arrows, left to right: light source, wavelength selector, sample in cuvette, detector.

A spectrophotometer shines a lamp through a wavelength selector, which picks out a single color of light. That light passes through the sample, which sits in a small, clear tube called a cuvette. A detector on the other side measures how much light made it through, and the display reports the result as percent transmittance or absorbance. Most school cuvettes are 1.00 cm wide, so the light always travels the same distance through the sample. A colorimeter works the same way but is simpler: instead of a wavelength selector, it shines one of a few fixed colors of light, usually from small LEDs, so you pick the color closest to the one the solution absorbs.

Transmittance and Absorbance

  • Transmittance (%T): % of light that passes through
  • Absorbance (A): light that is absorbed, no units
  • A goes up → %T goes down
  • 100% T means A = 0
Top: a pale pink cuvette lets a large arrow of light through, labeled high %T, low A. Bottom: a dark pink cuvette with many particles lets only a thin arrow of light through, labeled low %T, high A.

Transmittance and absorbance describe the same light from two sides. Transmittance is the percent of the light that passes through the sample. Absorbance tells how much of the light the sample took in, and it has no units. They move in opposite directions: a pale solution has a high %T and a low A, while a dark solution has a low %T and a high A. If 100% of the light gets through, nothing was absorbed and A = 0. Beer’s law uses absorbance, because absorbance is the value that grows in step with concentration.

The Blank

  • Cuvette with solvent only (water)
  • Set the instrument to A = 0 (100% T)
  • Removes light absorbed by the cuvette and solvent
  • Re-blank if you change the wavelength
A beam of light passes through a cuvette of clear water labeled Blank (water), and a display next to it reads A = 0.000.

The cuvette and the water also absorb and reflect a little light. To make sure the instrument only measures the solute, you first measure a blank: a cuvette filled with just the solvent, usually distilled water. You set the instrument to read A = 0 (100% T) with the blank inside. Every reading after that counts only the light absorbed by the dissolved particles. The blank absorbs differently at each wavelength, so blank the instrument again whenever you change the wavelength.

Choosing a Wavelength

  • Use λmax: the wavelength of maximum absorbance
  • Biggest change in A when c changes
  • Use the same wavelength for every reading
Absorbance spectrum of CoCl2(aq): absorbance on the y-axis and wavelength from 400 to 700 nm on the x-axis. The curve rises to one peak, marked with a dashed line labeled lambda max, about 510 nm.

A solution does not absorb every color equally. An absorbance spectrum shows how much light a substance absorbs at each wavelength, and the peak is called λmax (lambda max). Cobalt(II) chloride absorbs most strongly near 510 nm, which is why the class lab sets the spectrophotometer to 500 nm. At or near λmax, a small change in concentration makes the biggest change in absorbance, so the measurement is as sensitive as possible. Once you pick a wavelength, keep it the same for every standard and the unknown.

Beer’s Law Equation

  • A = εLc
  • A = absorbance (no units)
  • ε = absorption coefficient (M−1cm−1), depends on the substance and wavelength
  • L = path length (cm), usually 1.00 cm
  • c = concentration (M)
The equation A = εLc in a yellow box. Four cards below explain each symbol: A is absorbance, with no units. Epsilon is the absorption coefficient in inverse molar per centimeter, which depends on the substance and the wavelength. L is the path length in cm, the cuvette width, usually 1.00 cm. c is concentration in M. A note says some books write A = εbc.

Beer’s law connects absorbance to concentration. The Greek letter ε (epsilon) is the absorption coefficient, also called molar absorptivity. It measures how strongly one substance absorbs one wavelength of light, so it is a constant for that substance at that wavelength. L is the path length, the distance the light travels through the sample, which is the width of the cuvette. Some books write the same equation as A = εbc, using b for path length. The units cancel neatly: M−1cm−1 times cm times M leaves no units, which matches absorbance.

A Direct Proportion

  • Same substance, wavelength, and cuvette
  • A is directly proportional to c
  • Double c → double A
  • Graph is a straight line through (0, 0)
A table: 0.020 M has absorbance 0.096, 0.040 M has absorbance 0.192, and 0.080 M has absorbance 0.384. Red arrows labeled times 2 show that each time the concentration doubles, the absorbance doubles.

In a lab, you use the same substance, the same wavelength, and the same cuvette for every sample, so ε and L do not change. That leaves absorbance depending only on concentration: A is directly proportional to c. If you double the concentration, you double the number of particles in the light path, and the absorbance doubles too. A graph of absorbance versus concentration is a straight line that passes through the origin, because a solution with no solute absorbs no light.

Making a Standard Curve

  • Make standards of known c (by dilution)
  • Measure A for each standard
  • Graph A (y) vs. c (x)
  • Draw a best-fit line through (0, 0)
A standard curve: absorbance on the y-axis versus concentration in M on the x-axis. Four standards at 0.020, 0.040, 0.060 and 0.080 M lie on a straight blue best-fit line that starts at the origin.

In practice, chemists rarely look up ε. Instead they build a standard curve (also called a calibration curve). First, make several standards, solutions with known concentrations, usually by diluting a stock solution with M1V1 = M2V2. Measure the absorbance of each one at the same wavelength. Then graph absorbance on the y-axis against concentration on the x-axis and draw a straight best-fit line through the origin. Real data never lands perfectly on the line, so the best-fit line averages out small measurement errors.

Finding an Unknown

  • slope = A ÷ c (units: M−1)
  • slope = εL
  • cunknown = Aunknown ÷ slope
  • Or read it from the line
The same standard curve with a red unknown point on the line at absorbance 0.275. Dashed red lines run across to the absorbance axis and down to 0.050 M on the concentration axis.

The slope of the best-fit line is rise over run, or absorbance divided by concentration, so its units are M−1. Comparing y = mx to A = εLc shows that the slope equals εL. To find an unknown, measure its absorbance and divide by the slope. In the graph, the slope is 0.440 ÷ 0.0800 M = 5.50 M−1, so an unknown with A = 0.275 has c = 0.275 ÷ 5.50 M−1 = 0.0500 M. You can also find the unknown’s absorbance on the y-axis, go across to the line, and read straight down to the concentration. When you calculate a slope, use two points on the line itself, not two data points that sit off of it.

Sources of Error

MistakeMeasured ACalculated c
Fingerprints or smudges on cuvettetoo hightoo high
Water drops left in cuvette (sample diluted)too lowtoo low
Air bubbles in the light pathtoo hightoo high

Anything that blocks light makes the instrument think the sample absorbed it. Fingerprints, smudges, scratches, and air bubbles scatter or absorb light, so less light reaches the detector and the absorbance reads too high. Since c = A ÷ slope, a high absorbance gives a calculated concentration that is too high. Water left inside a cuvette does the opposite: it dilutes the sample, so fewer particles are in the light path, the absorbance reads too low, and the calculated concentration is too low. To avoid these errors, rinse the cuvette with the sample, wipe the outside with a KimWipe, and hold it by the top.

↑ Back to top


Worked Examples

Try each problem on your own first, then open the solution to check your work.

Example 1: Solving for Concentration

A CoCl2 solution has an absorbance of 0.384 in a 1.00 cm cuvette. The absorption coefficient is 4.80 M−1cm−1. What is the concentration?

Show the solution
  1. Known: A = 0.384, ε = 4.80 M−1cm−1, L = 1.00 cm
  2. Unknown: c, the concentration
  3. Rearrange: c = A ÷ (εL)
  4. Plug in: c = 0.384 ÷ (4.80 M−1cm−1 × 1.00 cm). The cm cancel, and dividing by M−1 leaves M.
  5. Calculate: 4.80 × 1.00 = 4.80, and 0.384 ÷ 4.80 = 0.0800 (three sig figs)
  6. Check: 4.80 × 1.00 × 0.0800 = 0.384, which matches the absorbance.
  7. Answer: c = 0.0800 M

Example 2: Using a Proportion

A 0.0400 M solution has an absorbance of 0.192. Using the same cuvette and wavelength, a second solution of the same substance has an absorbance of 0.336. What is its concentration?

Show the solution
  1. Known: c1 = 0.0400 M, A1 = 0.192, A2 = 0.336
  2. Unknown: c2, the second concentration
  3. Key idea: same substance, wavelength, and cuvette, so A ÷ c stays the same
  4. Find A ÷ c: 0.192 ÷ 0.0400 M = 4.80 M−1
  5. Calculate: c2 = 0.336 ÷ 4.80 M−1 = 0.0700 M (three sig figs)
  6. Check: A went up 0.336 ÷ 0.192 = 1.75 times, and 0.0400 M × 1.75 = 0.0700 M.
  7. Answer: c = 0.0700 M

Example 3: Reading a Standard Curve

A student made a standard curve for CoCl2. The best-fit line passes through the origin and these standards. (a) Calculate the slope. (b) An unknown has A = 0.360. What is its concentration?

Concentration (M)Absorbance
0.0300.144
0.0600.288
0.0900.432
0.1200.576
Show the solution
  1. Pick two points on the line: (0, 0) and (0.120 M, 0.576)
  2. Slope: rise ÷ run = (0.576 − 0) ÷ (0.120 M − 0) = 4.80 M−1
  3. Rearrange: c = A ÷ slope
  4. Plug in: c = 0.360 ÷ 4.80 M−1
  5. Calculate: 0.360 ÷ 4.80 = 0.0750 (three sig figs)
  6. Check: 0.360 is between 0.288 and 0.432, and 0.0750 M is between 0.060 M and 0.090 M.
  7. Answer: (a) slope = 4.80 M−1; (b) c = 0.0750 M

Example 4: Error Analysis

A student leaves fingerprints on the outside of the cuvette holding the unknown. How does this affect the measured absorbance and the calculated concentration? Justify your answer.

Show the solution
  1. What the fingerprints do: they absorb and scatter some of the light.
  2. Effect on the detector: less light reaches the detector, and the instrument counts all missing light as absorbed by the sample.
  3. Effect on A: the measured absorbance is too high.
  4. Effect on c: c = A ÷ slope, so a high A gives a calculated concentration that is too high.
  5. Answer: absorbance too high, concentration too high

↑ Back to top


Practice

Practice with the Beer’s Law Spectrophotometer Lab

Measure standards with a virtual spectrophotometer, build a calibration curve, and use it to find the concentration of an unknown.

Lab worksheet: PDF | Answer key: PDF

↑ Back to top


Lesson Slides

↑ Back to top


Big Ideas

  • Darker solutions absorb more light
  • Beer’s law: A = εLc
  • Blank first, then measure at λmax
  • Standard curve: slope = A ÷ c, so c = A ÷ slope
  • Smudges raise A; leftover water lowers A

Back to the essential question: A darker solution has more solute particles in the light path, so it absorbs more light. A spectrophotometer turns that color into a number, the absorbance, and Beer’s law shows that absorbance is directly proportional to concentration. By measuring standards and graphing a standard curve, you can turn the absorbance of any unknown into its concentration. Next up: Solution Stoichiometry.